Appendix: "Differential Lead Retention in
Zircons"
The most important results came from the thermal ionization experiments. The thermal ionization
mass spectrometer used in this work is similar to others described previously (3). It has a single magnet with 90° deflection and a 30-cm central radius of
curvature. It is equipped with a pulse-counting detection system to allow complete isotopic
analyses to be made on small quantities(<1 ng) of suitable elements ionized from a single
filament. The filaments, made of V-shaped rhenium foil 0.64 cm long and 0.08 cm deep (4), were baked out at 2000°C before loading the zircons. Ions are formed by
resistive heating of the filament; typical temperatures for this work were 1400° to 1470°C
(uncorrected pyrometer readings).
Previous work done to develop a technique for analyzing small lead samples led to the use of
silica gel to enhance ionization efficiency (5). Because individual zircons
are chemically somewhat similar to silica, we decided to try to analyze lead from individual
zircons loaded directly on the rhenium filament. Such a technique would have several advantages
over traditional methods: contamination would be essentially eliminated because no chemical
separation would be required and, since the zircons are small (~ 50 μm in diameter), they would
provide an approximate point source of ions, which is known to optimize ion-optical conditions in
the mass spectrometer (6).
Test experiments with zircons from other localities (7) were uniformly
successful; ion signals were observed at masses (m) 206, 207, and 208 which could definitely
be ascribed to Pb isotopes. To help ensure that we were at the correct ion lens conditions, we
focused on the 138BaO+ peak (the zircons contained some Ba), which was
reasonably intense at 1200°C. Surficial residues left on the zircons after the acetone wash
burned off before the operating temperature of 1450°C, where the lead signal was measured.
Great care had to be exercised to avoid making the temperature too high; very rapid evaporation of
the lead occurred only a little above the operating temperature. Typical count rates were 100 to
3000 counts per second for 206Pb+. Traces of thallium (m = 203 and
205) were sometimes observed, but burned out more rapidly than the lead. Other than thallium, lead
gave the only substantive peaks in the range m = 202 to 210. There was, however, a general
background generated by the sample; chemically unseparated samples such as these zircons almost
always yield such backgrounds. This background has little effect on the 206, 207, and 208 peaks,
but made precise measurement of the 204Pb signal, which was very small, impossible. For
example, in an analysis typical of these experiments, 1.6 × l05 counts from
206Pb were collected; the background correction was about 40 counts and, after correction, 18
counts remained at mass 204. Although these counts are listed as 204Pb counts in Table
2, more work is needed to determine how much may be uncompensated background.
Table 2 shows the results of our mass analyses of filaments loaded with single and multiple
zircons from five granite cores. The range of 206Pb/208Pb values reflects the
fact that this ratio varied from one group of zircons to another, and sometimes varied during
measurements on a single zircon. These variations are not surprising in view of the ion microprobe
analyses, which showed significant U/Th variations at different points on a single zircon
(232Th decays to 208Pb and 238U decays to 206Pb). These
variable 206Pb/208Pb ratios do not furnish any direct information on
differential Pb retention in these zircons. For that purpose, it is generally accepted that the
Radiogenic 206Pb/207Pb ratios derived from 238U/235U
decay are more specific. We note that Zartman's (8) isotopic
measurements of Pb, which was chemically extracted from zircons taken from the GT-2 core at 2900 m,
yield an adjusted 206Pb/207Pb ratio (9) that
approximates our ratios.
In a conventional chemical extraction of lead from zircons, the lead measured in the mass
analysis is considered to be a combination of radiogenic lead (from U and Th decay) and
nonradiogenic lead (from common lead contamination and from some initial lead in the zircon). The
radiogenic component is obtained by subtracting out a nonradiogenic component proportional to the
amount of 204Pb. In our experiments, however, the direct loading procedure virtually
eliminated the common lead contamination, and we circumvented the need to make adjustments for
initial lead in the zircons by accepting only analyses (10) showing a
ratio of 204Pb to total Pb of less than 2 × 10−3. Thus the
206Pb/207Pb ratios shown in Table 2 represent highly radiogenic lead and hence are
potential indicators of Pb retention.
We consider that the most important observations on the data in Table 2 are: (i) the fact that
the 206Pb/207Pb ratios on single zircons closely approximate the ratio
obtained when a group of similar zircons was loaded simultaneously on a single filament, (ii) the
relative uniformity of the 206Pb/207Pb ratios for zircons from all depths,
and (iii) the fact that the total number of Pb counts per zircon (the counts in column 4 of Table 2
divided by the product of columns 2 and 3) shows no systematic decrease with depth, as would be
expected if differential Pb loss had occurred at higher temperatures. Taken together, items (ii)
and (iii) provide strong evidence for high Pb retention in zircons even for a prolonged period in
an environment at an elevated temperature. These results have possible implications for long-term
nuclear waste disposal.
Zircon
depth (m) |
Filaments
analyzed |
Average
zircons
per
filament |
Total Pb
counts |
Counts
of 204Pb |
204Pb/
total Pb |
Average
206Pb/
208Pb |
Range
206Pb/
208Pb |
960 |
4 |
~ 10 |
1.2 × 106 |
235 |
2 × 10−4 |
9.6 ± 0.3 |
6.5-9.2 |
960 |
4 |
1 |
1.3 × 105 |
35 |
2.7 × 10−4 |
9.9 ± 0.4 |
5.8-14 |
2170 |
3 |
~ 5 |
8.9 × 105 |
269 |
3 × 10−4 |
10.0 ± 0.4 |
6.4-12.4 |
2900 |
3 |
~ 4 |
4.1 × 105 |
114 |
2.8 × 10−4 |
11.2 ± 0.3 |
4-11.4 |
3930 |
2 |
~ 10 |
6.5 × 105 |
132 |
2 × 10−4 |
11.0 ± 0.4 |
5.9-8.7 |
3930 |
2 |
1 |
8.0 × 104 |
46 |
5.8 × 10−4 |
10.4 ± 0.1 |
3.1-6.9 |
4310 |
7 |
~ 10 |
5.6 × 104 |
1400 |
2.5 × 10−4 |
9.7 ± 0.6 |
3.4-9.8 |
4310 |
2 |
1 |
1.6 × 105 |
100 |
6 × 10−4 |
9.8 ± 0.4 |
4.5-10.7 |
Table 2. Results of thermal ionization mass measurements for zircons with a
204Pb/total Pb ratio of less than 2 × 10−3. The background correction
was taken from the 208.5 mass position; it was applied to the raw data to obtain the isotopic
abundances, which were used to compute the isotopic ratios. Standard deviations are listed with the
Pb isotopic ratios.
|
|